ASTM D – Standard Test Method for Dielectric Breakdown Voltage of Insulating Oils of Petroleum Origin. Using VDE Electrodes. ▫ IEC Insulating. Page 1. Page 2. Page 3. Page 4. Find the most up-to-date version of ASTM D at Engineering
Author: | Gardall Kagatilar |
Country: | Puerto Rico |
Language: | English (Spanish) |
Genre: | Literature |
Published (Last): | 16 March 2009 |
Pages: | 405 |
PDF File Size: | 12.22 Mb |
ePub File Size: | 20.14 Mb |
ISBN: | 977-7-76179-530-2 |
Downloads: | 4208 |
Price: | Free* [*Free Regsitration Required] |
Uploader: | Nar |
Avoid touching the electrodes or the inside of the cell after cleaning. Link to Active This link will always route to the current Active aetm of the standard.
In order to preclude stirring air with the sample, provide the cell with a cover or baffle that will effectively prevent air from contacting the circulating liquid. Email Us or call toll-free: Referenced Documents purchase separately The documents listed below are referenced within the subject standard but are not provided as part of the standard.
High Voltage Inc TCD-5 Test Cell 60KV – ASTM D test specs – Mitchell Instrument Company
Reinstall the electrodes in the test cup and adjust spacing and clean in accordance with 7. The sample shall be taken in a dry, clean, non-permeable bottle. It is important to avoid touching the electrodes or the cleaned gage with aatm fingers or with portions of the tissue paper or chamois that have been in contact d18116 the hands.
It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Up to 75kV rms symmetrical Voltage measurement accuracy: Determination of the dielectric breakdown voltage of insulating oils provides an early detection method for any reduction in the insulating properties. ASTM Method being used? However, a high dielectric breakdown voltage does not necessarily indicate the absence of all contaminants; it may merely indicate that the concentrations of contaminants that are present in the liquid between d18816 electrodes are not large enough to deleteriously affect the average breakdown voltage of the liquid when tested by this test method see Appendix X1.
Phenix Technologies – Phenix Technologies TC/VDE Accessory For Astm D – John Morris
Clearance from the electrodes to all sides, bottom, cover or baffle, and any part of the stirring device is at least USB memory stick Operating Temperature: Petroleum Products and Biofuel lab apparatus baths, calorimeters, color measurement, density meters, flash point, hydrogenators, oil testing, penetrometers, surface tension, vapor pressure, viscometers and rheometers, At the beginning of each days testing, clean according to 7. After careful inspection, any electrodes from which pitting cannot be removed by light buffing should be discarded, as more refinishing would destroy the electrode contour and dimensions shown in Fig.
The most important requirement of an insulating oil is a high dielectric strength.
Measurement of oil temperature: Provide the test cell with a motor-driven two-bladed impeller and drive shaft, constructed asm a material having high dielectric strength. Contact with us now Your mail: It has been found to be especially useful in diagnostic and laboratory investigations of the dielectric breakdown strength of insulating liquid in insulating systems.
It is recommended that the circuit design limit the disruptive current duration and magnitude to low values that will minimize damage to the electrodes and limit formation of non-soluble materials resulting from the breakdown, but consistent with the requirements of 4.
If the breakdown voltage is in the expected range for this conditioned oil, the cell is considered properly prepared for testing other samples. Lab Equipment for Sale: IEEE Standard 4 is required during the manufacturing of the test apparatus and utilized during calibration of the equipment.
BAUR Report Manager
Automatic equipment should be used to control the voltage rate of rise because of the difficulty of maintaining a uniform voltage rise manually.
The dielectric breakdown voltage serves to indicate the presence of contaminating agents such as water, dirt, cellulosic fibers, or conducting particles in the liquid, one or more of which may be present in significant concentrations when low breakdown voltages are obtained.
Adjustment and Care of Electrodes and Test Cell 7. This test method should be used as recommended by professional organization standards such as IEEE C Refer to Terminology D for definitions used in this test method.
Automatic Dielectric Oil Tester 100 kv
Pharmaceutical laboratory apparatus stability chambers, tablet disintegration, tablet dissolution, tablet hardness, V-blenders, The voltage in the test circuit should be stable enough to be unaffected by asgm current flowing in the capacitive and resistive paths of the test circuit. Alternatively, the cell may be stored empty in a dust-free cabinet. Conduct a voltage breakdown test on a specimen of this oil in the manner specified in this test method. If a dielectric breakdown does not occur during any of the consecutive breakdown tests using the 2 mm spacing or the sample is not adequate for the 2 mm spacing test cell a d11816 more or less 0.
The test atsm shall be designed to permit easy removal of the electrodes for cleaning and polishing, verification that the shape is within the specified tolerance, and to permit easy adjustment of the gap spacing. What are your requirements.
Looking for a specific close-up? Mount the electrodes rigidly from opposite sides with the spacing axially centered within more or less 1 mm. Type text as it appears:: Calibration s shall be traceable to national standards and calibration should be verified annually or more often to ensure accuracy requirements.
Asrm equipment should produce a straight-line voltage-time curve over the operating range of the equipment.
When not in use, keep the cell filled with oil that meets the requirements of Specification D of the type normally tested. So that the breakdown may be observed, transparent materials are desirable, but not essential. Product Description The majority of high-voltage transformers, cables, aetm, transducers, capacitors, and rectifiers use insulating oils for insulating electrically live parts and to carry off thermal energy.
Power by seonoco All Rights Reserved: A lower than anticipated value is considered as evidence of cell contamination; then repeat the cleaning and the breakdown test with clean dry oil.